Solution

Common Problems and Solutions in ThermoCouple Calibration

2023-1-10 This article is read 3551 Times

ThermoCouple calibration system is mainly composed of intelligent multichannel thermometer, temperature calibration system software, precision liquid calibration bath, thermocouple calibration furnace, etc.It is mainly used for automatic calibration and verification of various thermocouples, thermal resistances, various thermometers, etc. The entire calibration process is automatically completed under the control of the temperature calibration/verification system software. When working, what are the common problems and solutions in thermocouple calibration? The editor briefly introduces to you the solution to the unqualified thermocouple.

ThermoCouple calibration system

The new nickel-chromium-nickel-silicon thermocouple (type K) is unqualified in the calibration

Solution:

During the verification work, I encountered a newly purchased K-type base metal thermocouple, which was judged as unqualified after the calibration. After analysis, it may be that the manufacturer did not anneal all the wires used to make the thermocouple. The solution is to carry out verification after the annealing procedure. 

After the thermocouple calibration program is started, the temperature of the reference junction displays "0.00" or prompts "the reference junction is abnormal"

Solutions: 

(1) This situation usually occurs after the digital multimeter is sent for inspection and reinstalled. Check whether the communication settings of the digital multimeter are correct, whether the positions of the front and rear keys are set correctly, and whether the connection with the scanner is correct

(2) You can also use the drop-down items "Scanner Communication Test" and "Digital Meter Communication Test" in the menu bar of the program interface to find the problem.

During the verification, the temperature of the thermocouple calibration furnace cannot be stabilized to the specified index for a long time.

Solution:

(1) Check whether the temperature control PID parameters of thermocouple calibration furnace in the verification software have been changed; 

(2) If the millivolt value displayed by the digital multimeter jumps at this time, there is probably interference, check whether the outer shell of the verification furnace is well grounded; 

( 3) Check whether the wiring circuits collide; 

(4) It may also be caused by poor contact of the power circuit of thermocouple calibration furnace, find out the fault point and deal with it.

Common Problems and Solutions in ThermoCouple Calibration
  • Online message

  • Your name
    Your title must be filled
  • Cellphone number
    Mobile phone number is required
  • whatsapp/wechat
    Optional
  • Type of consultation
    Required, please select the type of consultation
  • Consultation content
    Required, what is the content you want to consult?
  • Verification code: Please fill in the calculation result.
Previous:Thermocouple Calibration Furnace Pricenext:New Product | DTMC-mK301G mK high-precision thermometer updated

Copyright? Tai'an Dearto Automation Instrument Co., Ltd.

Chinese

绥中县| 盐亭县| 固始县| 化德县| 和硕县| 武隆县| 田东县| 溧水县| 剑阁县| 中宁县| 梓潼县| 宁蒗| 许昌县| 九寨沟县| 视频| 图们市| 银川市| 宣恩县| 平江县| 东方市| 咸宁市| 巢湖市| 绍兴市| 苍南县| 齐齐哈尔市| 吴忠市| 平昌县| 黄梅县| 四平市| 梁平县| 青阳县| 尼勒克县| 延川县| 武夷山市| 香港 | 桂平市| 松潘县| 韶关市| 肃宁县| 塔城市| 赣州市|