Solution

FAQs When Starting the Calibration Bath?

2022-12-5 This article is read 3034 Times

After using the calibration bath for several years, sometimes there will be problems such as the calibration bath does not heat up and the motor does not work. What should I do at this time?

calibration bath

1. Problem: The oil bath does not heat up, or the phenomenon of tripping when starting up.

Reason: The above-mentioned problem occurs in the calibration bath, and the heating rod may be leaking or damaged.

Solution

Open the terminal of the device, take a digital multimeter, and measure the resistance value of the device. If there is no resistance, the heating rod is bad. Need to contact Dearto to resend heater.

2. Problem: Turn on the stirring button of the calibration bath, and the motor does not work.

Reason: There are three reasons for this phenomenon:

1. The starting capacitor of the stirring motor is damaged;

2. The stirring motor is damaged;

3. The stirring switch is damaged.

Solution

1. If you start the motor by hand when the stirring power is turned on, if it can start, the starting capacitor is damaged, just replace the capacitor;

2. If the motor is started by hand when the stirring power is turned on, if it cannot be started, it means that the stirring motor is damaged, and the stirring motor can be replaced;

3. If the motor is started by hand when the stirring power is turned on, if it cannot be started, it means that the stirring motor is damaged or the stirring switch is damaged. After the power is turned off, check whether the switch is turned on. If the conduction does not work, the stirring switch is damaged.


FAQs When Starting the Calibration Bath?
  • Online message

  • Your name
    Your title must be filled
  • Cellphone number
    Mobile phone number is required
  • whatsapp/wechat
    Optional
  • Type of consultation
    Required, please select the type of consultation
  • Consultation content
    Required, what is the content you want to consult?
  • Verification code: Please fill in the calculation result.
Previous:What is Temperature Calibration?next:New Product | DTMC-mK301G mK high-precision thermometer updated

Copyright? Tai'an Dearto Automation Instrument Co., Ltd.

Chinese

扬州市| 庆元县| 长治县| 平塘县| 错那县| 汝城县| 共和县| 山丹县| 油尖旺区| 龙泉市| 曲沃县| 革吉县| 海南省| 凌源市| 永仁县| 嘉禾县| 昭觉县| 安达市| 陆川县| 郁南县| 比如县| 佛坪县| 南郑县| 洛川县| 荥经县| 班戈县| 广宁县| 高要市| 阿城市| 尖扎县| 沅陵县| 镶黄旗| 永寿县| 江西省| 彭阳县| 祁阳县| 凌源市| 桂东县| 自贡市| 黎城县| 利辛县|