a东北肥胖男人肛交HD,97青娱乐青娱乐色婷婷,男生和女生透鸡鸡的网站,亚洲女同志亚洲女同女同

News

Ultimate Knowledge | Calibration method of temperature and humidity calibration chamber

2024-2-5 This article is read 2393 Times

DTLH copywriting.jpg

1. Sensor location

Before calibration, the effective working area shall  be  determined  according to the  operation manual of the temperature and humidity chamber or the user   requirements.Multi-channel temperature and humidity measuring device is used to calibrate the temperature and humidity in the chamber. The temperature sensor and humidity sensor in the device adopt horizontal layered layout. The location from the side of the effective working area is 8 cm. The lower location is 5 cm above the bottom level of the effective working area, and the upper location is 20 cm below the top level of the

effective working area, as shown in Figure 1.

Temperature and humidity chamber with the effective working area height,h,less than 50 cm,

locate 5 points [see Figure la)

Temperature and humidity chamber with the effective working area height ,h, between 50 c m

and 100 c m, locate 9 points [see Figure lb)

Temperature and humidity chamber with the height of the effective working area,h,exceeding

100 cm, locate 13 points [see Figure lc)].

a)   h <50 cm distribution plot

b) 50 cm ≤ h ≤100 cm  distribution plot


c) h> 100 cm distribution plot

Figure 1 Schematic diagram of the sensor spot placement

For  the  temperature and humidity  chamber  with  input cold mirror dew point probe in  the working room, the distribution position of the sensor should be adjusted appropriately, so that the distance between the sensor in the upstream of the dew point probe and the air flow in the chamber is not less than 5 cm, and the distance between the sensor downstream of the dew point probe and the

dew point probe is not less than 20 cm.

For the temperature and humidity chamber with airflow blind area in the effective working area, a distribution point can be added in the blind area.


2. Temperature calibration

2. 1 Temperature calibration point

Temperature calibration points are generally 15 °C, 20 °C, and 30 °C, and other calibration points

can also be selected according to user requirements.During the calibration, the temperature shall be conducted in the order of low to high, and the temperature deviation at each calibration point shall be within  ±0.5 °C (whichever is the center point temperature).If the humidity function is turned on, the humidity  set  value  is  generally  about  60%RH.The  calibration  items  are  generally  conducted  under no-load conditions.

2. 2 Temperature calibration steps

After   the   temperature   and    humidity   chamber   reaches   the   specified    stability   time,   the temperature  of  each  point  is  recorded.  The  recording  interval  of  the  center  point  is  20s,  and  the recording interval of the other points is 2 min.Center points recorded 91 sets of data and other points recorded 16 sets of data.20s interval data use T ’i0  (i =  1,2, ??? ,91) indicates that 2 min interval data were used with Tnj (n=l, 2, ??? , 16; j =0,1, ??? , m-1), where m is the number of points, i and n represents the serial  number  of time  (i  time  interval  is  20s,n  time  interval  is  2  min),  j  representing  the  serial number of point position.If the temperature and humidity in the chamber can be determined to be

balanced before the specified stability time, it can also be recorded in advance.

The temperature stability time shall be subject to the instructions, if not given in the instructions, the following principles: if the temperature change does not exceed 10 °C, the stability time from the temperature and humidity setting to the start recording data is 90 min; if the temperature change does not exceed 5 °C, the stability time is 60 min.The stabilization times of each calibration point are shown

in Table 1.

Table 1 Stability Time



Ultimate Knowledge | Calibration method of temperature and humidity calibration chamber
  • Online message

  • Your name
    Your title must be filled
  • Cellphone number
    Mobile phone number is required
  • whatsapp/wechat
    Optional
  • Type of consultation
    Required, please select the type of consultation
  • Consultation content
    Required, what is the content you want to consult?
  • Verification code: Please fill in the calculation result.
Previous:New Product | 180C to 670C temperature calibration intelligent precision thermostatic salt bathnext:New Product | DTMC-mK301G mK high-precision thermometer updated

Copyright? Tai'an Dearto Automation Instrument Co., Ltd.

Chinese

石嘴山市| 中方县| 河东区| 博野县| 资阳市| 望都县| 资中县| 随州市| 邢台市| 吴川市| 海城市| 重庆市| 黑龙江省| 和平区| 江都市| 黔江区| 永宁县| 洛阳市| 镇雄县| 尼木县| 武邑县| 潢川县| 威远县| 兴业县| 曲靖市| 仁布县| 天峨县| 林州市| 图们市| 武平县| 漳平市| 东光县| 建水县| 汤原县| 灵武市| 上饶市| 哈巴河县| 丹巴县| 明水县| 蕉岭县| 乌什县|