DEARTO-Professional manufacturer of temperature and humidity measuring instruments
Tai'an Detu Automation Instrument Co., Ltd.

0538-5089056

Solution Current position:Home > Solution

Common Problems and Solutions in ThermoCouple Calibration

Release time:2023-1-10    This article is read 3549 Times

ThermoCouple calibration system is mainly composed of intelligent multichannel thermometer, temperature calibration system software, precision liquid calibration bath, thermocouple calibration furnace, etc.It is mainly used for automatic calibration and verification of various thermocouples, thermal resistances, various thermometers, etc. The entire calibration process is automatically completed under the control of the temperature calibration/verification system software. When working, what are the common problems and solutions in thermocouple calibration? The editor briefly introduces to you the solution to the unqualified thermocouple.

ThermoCouple calibration system

The new nickel-chromium-nickel-silicon thermocouple (type K) is unqualified in the calibration

Solution:

During the verification work, I encountered a newly purchased K-type base metal thermocouple, which was judged as unqualified after the calibration. After analysis, it may be that the manufacturer did not anneal all the wires used to make the thermocouple. The solution is to carry out verification after the annealing procedure. 

After the thermocouple calibration program is started, the temperature of the reference junction displays "0.00" or prompts "the reference junction is abnormal"

Solutions: 

(1) This situation usually occurs after the digital multimeter is sent for inspection and reinstalled. Check whether the communication settings of the digital multimeter are correct, whether the positions of the front and rear keys are set correctly, and whether the connection with the scanner is correct

(2) You can also use the drop-down items "Scanner Communication Test" and "Digital Meter Communication Test" in the menu bar of the program interface to find the problem.

During the verification, the temperature of the thermocouple calibration furnace cannot be stabilized to the specified index for a long time.

Solution:

(1) Check whether the temperature control PID parameters of thermocouple calibration furnace in the verification software have been changed; 

(2) If the millivolt value displayed by the digital multimeter jumps at this time, there is probably interference, check whether the outer shell of the verification furnace is well grounded; 

( 3) Check whether the wiring circuits collide; 

(4) It may also be caused by poor contact of the power circuit of thermocouple calibration furnace, find out the fault point and deal with it.

Common Problems and Solutions in ThermoCouple Calibration
  • Online message

  • Your name
    Your title must be filled
  • Cellphone number
    Mobile phone number is required
  • whatsapp/wechat
    Optional
  • Type of consultation
    Required, please select the type of consultation
  • Consultation content
    Required, what is the content you want to consult?
  • Verification code: Please fill in the calculation result.
Previous:Thermocouple Calibration Furnace Pricenext:New Product | DTMC-mK301G mK high-precision thermometer updated

Contact us

DEARTO QR codeHeadquarter Address Tai'an High-tech Development Zone,  Shandong Province 
Telephon :+86 0538-5089056  +86 13953856217 
Fax : +86 0538-5059718 
After sale:+86 0538-5050959
Website : www.auto-surfer.com 
E-mail   :  tadtzdh@dearto.cn

Copyright? Tai'an Detu Automation Instrument Co., Ltd.             Lu ICP 14024757-1           Business license inquiry

衡山县| 三门峡市| 托里县| 泸水县| 铜梁县| 德州市| 册亨县| 尉氏县| 三原县| 呼和浩特市| 隆安县| 达孜县| 西和县| 交口县| 木兰县| 三原县| 汝南县| 抚州市| 南陵县| 乾安县| 尉犁县| 溧阳市| 金沙县| 阿拉善左旗| 泰顺县| 永胜县| 伽师县| 灵武市| 陇西县| 铜陵市| 平安县| 临潭县| 澄江县| 茌平县| 平和县| 灵丘县| 盈江县| 昌图县| 双流县| 三台县| 达日县|