Common Problems and Solutions in ThermoCouple Calibration

Release time:2023-1-10    This article is read 3547 Times

ThermoCouple calibration system is mainly composed of intelligent multichannel thermometer, temperature calibration system software, precision liquid calibration bath, thermocouple calibration furnace, etc.It is mainly used for automatic calibration and verification of various thermocouples, thermal resistances, various thermometers, etc. The entire calibration process is automatically completed under the control of the temperature calibration/verification system software. When working, what are the common problems and solutions in thermocouple calibration? The editor briefly introduces to you the solution to the unqualified thermocouple.

ThermoCouple calibration system

The new nickel-chromium-nickel-silicon thermocouple (type K) is unqualified in the calibration

Solution:

During the verification work, I encountered a newly purchased K-type base metal thermocouple, which was judged as unqualified after the calibration. After analysis, it may be that the manufacturer did not anneal all the wires used to make the thermocouple. The solution is to carry out verification after the annealing procedure. 

After the thermocouple calibration program is started, the temperature of the reference junction displays "0.00" or prompts "the reference junction is abnormal"

Solutions: 

(1) This situation usually occurs after the digital multimeter is sent for inspection and reinstalled. Check whether the communication settings of the digital multimeter are correct, whether the positions of the front and rear keys are set correctly, and whether the connection with the scanner is correct

(2) You can also use the drop-down items "Scanner Communication Test" and "Digital Meter Communication Test" in the menu bar of the program interface to find the problem.

During the verification, the temperature of the thermocouple calibration furnace cannot be stabilized to the specified index for a long time.

Solution:

(1) Check whether the temperature control PID parameters of thermocouple calibration furnace in the verification software have been changed; 

(2) If the millivolt value displayed by the digital multimeter jumps at this time, there is probably interference, check whether the outer shell of the verification furnace is well grounded; 

( 3) Check whether the wiring circuits collide; 

(4) It may also be caused by poor contact of the power circuit of thermocouple calibration furnace, find out the fault point and deal with it.

Common Problems and Solutions in ThermoCouple Calibration
  • Online message

  • Your name
    Your title must be filled
  • Cellphone number
    Mobile phone number is required
  • whatsapp/wechat
    Optional
  • Type of consultation
    Required, please select the type of consultation
  • Consultation content
    Required, what is the content you want to consult?
  • Verification code: Please fill in the calculation result.
Previous:Thermocouple Calibration Furnace Pricenext:New Product | DTMC-mK301G mK high-precision thermometer updated

Contact us

DEARTO QR codeHeadquarter Address Tai'an High-tech Development Zone,  Shandong Province 
Telephon :+86 0538-5089056  +86 13953856217 
Fax : +86 0538-5059718 
After sale:+86 0538-5050959
Website : www.auto-surfer.com 
E-mail   :  tadtzdh@dearto.cn

Copyright? Tai'an Detu Automation Instrument Co., Ltd.             Lu ICP 14024757-1           Business license inquiry

奎屯市| 广河县| 通化县| 哈尔滨市| 庆城县| 万荣县| 城口县| 二手房| 吉木萨尔县| 黔西县| 沁源县| 通渭县| 滨州市| 丹巴县| 鸡西市| 资源县| 嘉黎县| 长白| 左权县| 漳浦县| 松潘县| 云南省| 盐津县| 稷山县| 宣化县| 嘉善县| 松原市| 台东市| 措勤县| 宁阳县| 新泰市| 吕梁市| 基隆市| 钟山县| 虞城县| 渭源县| 江北区| 邵阳市| 潼关县| 奉化市| 浦县|